FLIR GFx320
Support Center

FLIR GFx320

Go to Product Details Page

For Support
Please call 1-603-324-7600 or click here for support

For Service
Call 1-866-354-7911 or Submit RMA

Overview:
Recorded Webinar

Overview of Optical Gas Imaging Webinar

Watch now
Tech Note

Gases Detected and Minimum Detected leak rate (MDLR)

Read the Story
Technical Guidance

How can I stream video from my USB or Ethernet camera?

Learn more
Registration

I'm trying to register my FLIR product but I recieve an error that the serial number is not found. What is the reason?

Read the Story
Warranty & RMA

FLIR Standard Warranty

Learn more
Recorded Webinar

Techniques for Successful Gas Detection

Watch now
Compatibility

What cameras are supported in ResearchIR / ResearchIR MAX ?

Read the Story
Resources:
Datasheet: 16-0146-INS-OGI-GFx320-Datasheet
Download
Datasheet: 16-0146-ins-ogi-gfx320-datasheet-EMEA-EN-v3
Download
Certification: FLIR GF320 GF300 G300a EPA OOOOa Compliance Certificate
Download
User Manual: FLIR GFx3xx User Manual
Download
Brochure/Catalog: GFx320 Brochure
Download
Tech Note: Science Behind Optical Gas Imaging
Download
Specifications
Overview
Detector Type
FLIR Indium Antimonide (InSb)
Dynamic Range
14-bit
IR Resolution
320 x 240 pixels
Mounting
Standard, 1/4"-20
Spectral Range
3.2 – 3.4 µm
Zoom
1-8x continuous digital zoom
Hazardous Location Compliance
ATEX/IECEx, Ex ic nC op is IIC T4 Gc II 3 G ANSI/ISA-12.12.01-2013, Class I Div 2 CSA 22.2 No. 213, Class I Div 2
Environmental
Ambient temperature range
-20°C to 40°C (-4°F to 104°F) (Certification range for explosive atmospheres)
Encapsulation
IP 54 (IEC 60529)
Operating Temperature Range
–20°C to 50°C (–4°F to 122°F)
Shock & Vibration
25 g (IEC 60068-2-27) / 2 g (IEC 60068-2-6)
Standard Temperature Range
–20°C to 350°C (–4°F to 662°F)
Storage Temperature Range
–30°C to 60°C (–22°F to 140°F)
Imaging & Optical
Automatic Image Adjustment
Continuous/manual, linear, histogram
Camera f-number
f/1.5
Color palettes
Iron, Gray, Rainbow, Arctic, Lava, Rainbow HC
Detector Pitch
30 µm
Focus
Manual
Frame Rate [Full Window]
60 Hz
Global Positioning System [GPS]
Location data stored with every image
Image Modes
IR image, visual image, High Sensitivity Mode (HSM)
Sensor Cooling
Stirling Microcooler (FLIR MC-3)
Image Analysis
10 spotmeters, 5 boxes with max./min./average, profile, delta temperatures, emissivity & measurement corrections
Non-Radiometric IR Video
MPEG4 (up to 60 min./clip) to memory card
On Camera Display
Widescreen 800 x 480 pixel LCD Tiltable 800 × 480 pixel OLED viewfinder
Radiometric IR Video
15 Hz direct to memory card
Visual Image
3.2 MP from integrated visible camera Can be automatically associated with corresponding non-radiometric IR video
Visual Video
MPEG4 (25 min./clip) to memory card
Measurement & Analysis
Accuracy
±1°C (±1.8°F) for temperature range (0°C to 100°C, 32°F to 212°F) or ±2% of reading for temperature range (>100°C, >212°F)
NETD
<15 mK @ 30°C (86°F)
Power
Battery type
Rechargeable Li-ion battery
External Power Operation
AC adapter 90-260 VAC, 50/60 Hz or 12 VDC from a vehicle